 | 1. | Make sure the analyzer is turned off. |
 | 2. | Open the access door by pressing the top of the door as shown below. The bottom will open allowing you to remove the door. |
 | |
 Opening the Access Door |
 | 3. | If the old roll still contains paper, cut (or tear) the paper near the paper roll and pull the remnant forward through the top of the printer. Lift the old roll out of the paper roll holder. |
 |
CAUTION
Never pull the paper from inside the access door back down through the printer. To remove the paper from the printer, cut or tear the paper near the paper roll and pull the remnant forward through the top of the printer.
|
 | 4. | Remove and discard the first six inches of the new paper roll - this part of the roll usually contains glue and may jam the printer. Unroll an additional six inches of the new paper roll and cut the paper as shown below with scissors. |
 | |
 Cutting a New Print Roll |
 | 5. | Insert the end of the paper into the analyzer as shown below. |
 | |
 Inserting the Print Roll |